N2 - Introduction The aim of this ex vivo study was to evaluate the accuracy of the Raypex 5 (VDW, Munich, Germany) and Apex NRG XFR (Medic NRG Ltd, Tel Aviv, Israel) electronic apex locators (EALs) in determining the working length when compared with radiographs. Methods Twenty-five human single-rooted teeth were selected, and the access cavity was prepared. The working length (WL) was determined radiographically and electronically by using 2 EALs. The files were fixed at the WL, and the teeth were extracted. The apical 4 mm of each canal was trimmed to expose the file tip, and the samples were observed under a stereomicroscope. The distance from the file tip to the point 0.5 mm coronal to the anatomic apex was measured. The data were analyzed by using 1-way analysis of variance and the Tukey Honestly Significant Difference test. Results There was no significant difference between the Raypex 5 and the Apex NRG XFR devices with respect to their accuracy in determining the final WL. When compared with radiography, both the EALs had no significant difference. When comparing EALs and radiographic measurements with control measurements, accuracy results were found to be 20%, 36%, and 52% for the Raypex 5, Apex NRG XFR, and radiography, respectively. Overestimations of WL determination by the Raypex 5, Apex NRG XFR, and radiography were 4%, 0%, and 40%, respectively. Underestimations of WL determination by the Raypex 5, Apex NRG XFR, and radiography were 76%, 64%, and 8%, respectively. Conclusions Both the EALs had the same accuracy in determining the WL when compared with radiography.
AB - Introduction The aim of this ex vivo study was to evaluate the accuracy of the Raypex 5 (VDW, Munich, Germany) and Apex NRG XFR (Medic NRG Ltd, Tel Aviv, Israel) electronic apex locators (EALs) in determining the working length when compared with radiographs. Methods Twenty-five human single-rooted teeth were selected, and the access cavity was prepared. The working length (WL) was determined radiographically and electronically by using 2 EALs. The files were fixed at the WL, and the teeth were extracted. The apical 4 mm of each canal was trimmed to expose the file tip, and the samples were observed under a stereomicroscope. The distance from the file tip to the point 0.5 mm coronal to the anatomic apex was measured. The data were analyzed by using 1-way analysis of variance and the Tukey Honestly Significant Difference test. Results There was no significant difference between the Raypex 5 and the Apex NRG XFR devices with respect to their accuracy in determining the final WL. When compared with radiography, both the EALs had no significant difference. When comparing EALs and radiographic measurements with control measurements, accuracy results were found to be 20%, 36%, and 52% for the Raypex 5, Apex NRG XFR, and radiography, respectively. Overestimations of WL determination by the Raypex 5, Apex NRG XFR, and radiography were 4%, 0%, and 40%, respectively. Underestimations of WL determination by the Raypex 5, Apex NRG XFR, and radiography were 76%, 64%, and 8%, respectively. Conclusions Both the EALs had the same accuracy in determining the WL when compared with radiography.
UR - http://www.scopus.com/inward/record.url?scp=84921510587&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84921510587&partnerID=8YFLogxK
U2 - 10.1016/j.joen.2014.10.011
DO - 10.1016/j.joen.2014.10.011
M3 - Article
C2 - 25476973
AN - SCOPUS:84921510587
SN - 0099-2399
VL - 41
SP - 201
EP - 204
JO - Journal of Endodontics
JF - Journal of Endodontics
IS - 2
ER -