Solutions to kinking of the side stream carbon dioxide sampling line

Mike Young, Goneppanavar Umesh, Nilesh Kadam, Kaur Jasvinder

Research output: Contribution to journalArticle

4 Citations (Scopus)

Abstract

A sidestream gas analyzer samples gases from a sampling port in an adapter placed in a breathing circuit. Kinking of the sampling line of the sidestream capnogram is not an uncommon problem during head and neck surgeries. We have identified the main problem for kinking of the sampling line in head and neck surgeries to be its vertical origin from the connecting point of the breathing circuit. Angled attachment of the sampling line at its connection to the breathing circuit can minimize this problem of kinking significantly as has been shown by our study.

Original languageEnglish
Pages (from-to)221-222
Number of pages2
JournalJournal of Clinical Monitoring and Computing
Volume24
Issue number3
DOIs
Publication statusPublished - 01-06-2010
Externally publishedYes

Fingerprint

Carbon Dioxide
Respiration
Neck
Gases
Head

All Science Journal Classification (ASJC) codes

  • Medicine(all)
  • Health Informatics
  • Critical Care and Intensive Care Medicine
  • Anesthesiology and Pain Medicine

Cite this

Young, Mike ; Umesh, Goneppanavar ; Kadam, Nilesh ; Jasvinder, Kaur. / Solutions to kinking of the side stream carbon dioxide sampling line. In: Journal of Clinical Monitoring and Computing. 2010 ; Vol. 24, No. 3. pp. 221-222.
@article{f9027681d43d401d9829dbd03e808fea,
title = "Solutions to kinking of the side stream carbon dioxide sampling line",
abstract = "A sidestream gas analyzer samples gases from a sampling port in an adapter placed in a breathing circuit. Kinking of the sampling line of the sidestream capnogram is not an uncommon problem during head and neck surgeries. We have identified the main problem for kinking of the sampling line in head and neck surgeries to be its vertical origin from the connecting point of the breathing circuit. Angled attachment of the sampling line at its connection to the breathing circuit can minimize this problem of kinking significantly as has been shown by our study.",
author = "Mike Young and Goneppanavar Umesh and Nilesh Kadam and Kaur Jasvinder",
year = "2010",
month = "6",
day = "1",
doi = "10.1007/s10877-010-9237-1",
language = "English",
volume = "24",
pages = "221--222",
journal = "Journal of Clinical Monitoring and Computing",
issn = "1387-1307",
publisher = "Springer Netherlands",
number = "3",

}

Solutions to kinking of the side stream carbon dioxide sampling line. / Young, Mike; Umesh, Goneppanavar; Kadam, Nilesh; Jasvinder, Kaur.

In: Journal of Clinical Monitoring and Computing, Vol. 24, No. 3, 01.06.2010, p. 221-222.

Research output: Contribution to journalArticle

TY - JOUR

T1 - Solutions to kinking of the side stream carbon dioxide sampling line

AU - Young, Mike

AU - Umesh, Goneppanavar

AU - Kadam, Nilesh

AU - Jasvinder, Kaur

PY - 2010/6/1

Y1 - 2010/6/1

N2 - A sidestream gas analyzer samples gases from a sampling port in an adapter placed in a breathing circuit. Kinking of the sampling line of the sidestream capnogram is not an uncommon problem during head and neck surgeries. We have identified the main problem for kinking of the sampling line in head and neck surgeries to be its vertical origin from the connecting point of the breathing circuit. Angled attachment of the sampling line at its connection to the breathing circuit can minimize this problem of kinking significantly as has been shown by our study.

AB - A sidestream gas analyzer samples gases from a sampling port in an adapter placed in a breathing circuit. Kinking of the sampling line of the sidestream capnogram is not an uncommon problem during head and neck surgeries. We have identified the main problem for kinking of the sampling line in head and neck surgeries to be its vertical origin from the connecting point of the breathing circuit. Angled attachment of the sampling line at its connection to the breathing circuit can minimize this problem of kinking significantly as has been shown by our study.

UR - http://www.scopus.com/inward/record.url?scp=77956884833&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=77956884833&partnerID=8YFLogxK

U2 - 10.1007/s10877-010-9237-1

DO - 10.1007/s10877-010-9237-1

M3 - Article

C2 - 20559864

AN - SCOPUS:77956884833

VL - 24

SP - 221

EP - 222

JO - Journal of Clinical Monitoring and Computing

JF - Journal of Clinical Monitoring and Computing

SN - 1387-1307

IS - 3

ER -