Yet another cause for blocked sidestream capnogram - Beware of the non-threaded cap mount in heat and moisture exchangers

Goneppanavar Umesh, Kaur Jasvinder, Nanda Shetty

Research output: Contribution to journalArticle

Abstract

Heat and moisture exchangers (HME) are commonly used during general anaesthesia and intensive care of patients on mechanical ventilators. Some of the HME manufacturers provide HMEs with a Luer lock fitting for connecting side stream CO2 monitoring line, Luer lock cap, and a non-threaded cap mount. However, HMEs from different manufacturers and HMEs meant for use in children and for adults from the same manufacturer vary in the presence/absence of non-threaded cap mount. This can create confusion to the clinicians and can result in inadvertent connection of the CO2 monitoring line to the non-threaded cap mount resulting in blocked CO2 monitoring line and leak in the circuit. We caution all the anaesthesiologists and intensivists regarding this possibility while using HMEs from different manufacturers.

Original languageEnglish
Pages (from-to)207-208
Number of pages2
JournalJournal of Clinical Monitoring and Computing
Volume23
Issue number4
DOIs
Publication statusPublished - 09-06-2009
Externally publishedYes

Fingerprint

Hot Temperature
Mechanical Ventilators
Critical Care
General Anesthesia

All Science Journal Classification (ASJC) codes

  • Medicine(all)
  • Health Informatics
  • Critical Care and Intensive Care Medicine
  • Anesthesiology and Pain Medicine

Cite this

@article{f2b9f469e90b48ea8205a3e3f6034647,
title = "Yet another cause for blocked sidestream capnogram - Beware of the non-threaded cap mount in heat and moisture exchangers",
abstract = "Heat and moisture exchangers (HME) are commonly used during general anaesthesia and intensive care of patients on mechanical ventilators. Some of the HME manufacturers provide HMEs with a Luer lock fitting for connecting side stream CO2 monitoring line, Luer lock cap, and a non-threaded cap mount. However, HMEs from different manufacturers and HMEs meant for use in children and for adults from the same manufacturer vary in the presence/absence of non-threaded cap mount. This can create confusion to the clinicians and can result in inadvertent connection of the CO2 monitoring line to the non-threaded cap mount resulting in blocked CO2 monitoring line and leak in the circuit. We caution all the anaesthesiologists and intensivists regarding this possibility while using HMEs from different manufacturers.",
author = "Goneppanavar Umesh and Kaur Jasvinder and Nanda Shetty",
year = "2009",
month = "6",
day = "9",
doi = "10.1007/s10877-009-9183-y",
language = "English",
volume = "23",
pages = "207--208",
journal = "Journal of Clinical Monitoring and Computing",
issn = "1387-1307",
publisher = "Springer Netherlands",
number = "4",

}

Yet another cause for blocked sidestream capnogram - Beware of the non-threaded cap mount in heat and moisture exchangers. / Umesh, Goneppanavar; Jasvinder, Kaur; Shetty, Nanda.

In: Journal of Clinical Monitoring and Computing, Vol. 23, No. 4, 09.06.2009, p. 207-208.

Research output: Contribution to journalArticle

TY - JOUR

T1 - Yet another cause for blocked sidestream capnogram - Beware of the non-threaded cap mount in heat and moisture exchangers

AU - Umesh, Goneppanavar

AU - Jasvinder, Kaur

AU - Shetty, Nanda

PY - 2009/6/9

Y1 - 2009/6/9

N2 - Heat and moisture exchangers (HME) are commonly used during general anaesthesia and intensive care of patients on mechanical ventilators. Some of the HME manufacturers provide HMEs with a Luer lock fitting for connecting side stream CO2 monitoring line, Luer lock cap, and a non-threaded cap mount. However, HMEs from different manufacturers and HMEs meant for use in children and for adults from the same manufacturer vary in the presence/absence of non-threaded cap mount. This can create confusion to the clinicians and can result in inadvertent connection of the CO2 monitoring line to the non-threaded cap mount resulting in blocked CO2 monitoring line and leak in the circuit. We caution all the anaesthesiologists and intensivists regarding this possibility while using HMEs from different manufacturers.

AB - Heat and moisture exchangers (HME) are commonly used during general anaesthesia and intensive care of patients on mechanical ventilators. Some of the HME manufacturers provide HMEs with a Luer lock fitting for connecting side stream CO2 monitoring line, Luer lock cap, and a non-threaded cap mount. However, HMEs from different manufacturers and HMEs meant for use in children and for adults from the same manufacturer vary in the presence/absence of non-threaded cap mount. This can create confusion to the clinicians and can result in inadvertent connection of the CO2 monitoring line to the non-threaded cap mount resulting in blocked CO2 monitoring line and leak in the circuit. We caution all the anaesthesiologists and intensivists regarding this possibility while using HMEs from different manufacturers.

UR - http://www.scopus.com/inward/record.url?scp=68349127378&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=68349127378&partnerID=8YFLogxK

U2 - 10.1007/s10877-009-9183-y

DO - 10.1007/s10877-009-9183-y

M3 - Article

VL - 23

SP - 207

EP - 208

JO - Journal of Clinical Monitoring and Computing

JF - Journal of Clinical Monitoring and Computing

SN - 1387-1307

IS - 4

ER -